Infusion Set and/or Access Site problems :
Problem Corrective Action
Poor insulin absorption Check access site for lumps,
scarring or bruising
(hypertrophied areas). Move the
access site, if necessary.
Access site red, itchy, Deliver a Correction bolus to
irritated; bumps or lower your BG level. Move the
nodules at site access site, if necessary.
Air bubbles in Make sure all connections are
cartridge/infusion tight and not leaking (wrap tissue
set or leaks at around connectors to check for
connections moisture). Look for air bubbles in
cartridge and infusion set.
Disconnect infusion set from your
body and use the Fill tubing
feature to remove large air
bubbles, if needed.
Kinked tubing Inspect entire length of infusion
set for kinks, including the safety
loop. Make sure your clothing
and/or pump pouch are not
pinching/restricting tubing.
Change infusion set, if required.
Problem Corrective Action
Needle/cannula If needle/cannula is dislodged,
dislodged change infusion set and access
site.