Technical Information
108
Blockage (occlusion) alert:
18 psi ± 5psi
Bolus accuracy at set value of 0.1u:
± 15% (measurement error ± 10%)
Bolus accuracy at set value of 25u:
± 1.5% (measurement accuracy ± 0.5%)
Maximum volume infused under single fault condition:
Less than 2 units
Delivery rate during bolus:
Approximately 0.05 units per second minimum, 1.67
units per second maximum (based on bolus amount –
all standard boluses to be delivered within 1 minute)
Delivery rate during Fill tubing/Fill cannula:
Approximately 1 unit per second
Basal Delivery Interval (minimum):
3 minutes (approximately
1
/
20
basal rate)
Delivery specifications
Main Menu
Meal Bolus:
User can program and deliver a meal bolus. If displayed,
extended and combination bolus appear in menu. If
displayed, any custom boluses appear in menu
Delivery Specifications:
Programming/delivering a bolus (Standard, Extended,
and Combination):
Units - Insulin: 0 units to selected maximum bolus
Default: 0 units
Grams - Carb: 0 grams to selected maximum bolus ×
carb ratio
Default: 0 grams
Duration (Extended and Combination): 00:15 minutes to
8:00 hours in 00:15 minute increments
Default: 1:00 hour
Percent Immediate (Combination): 1% to 99% in 1%
increments
Default: 50%
Setup in Pump Settings / Personalize / Delivery / Meal
bolus menu
Type: Grams - Carb or Units - Insulin
Default: Units - Insulin
Maximum Bolus: 1.00 units to 25.00 units in 1.00 unit
increments
Default: 15.00 units
Zobrazit stránku 119