Important Information
5
PROOF (LR #3557), 2002-11-25 «5118-51A Cozmo User Manual»
lems, remove the pump from use.
☞
The pump may be damaged if stored in environments where
temperatures are lower than –4˚F (–20˚C) or higher than
140˚F (60˚C), when relative humidity (non-condensing)
exceeds 90%, and when atmospheric pressure is lower than
10.2 PSI (70 kPa) (10,000 feet above sea level) or higher
than 15.4 PSI (106 kPa).
☞
Do not use NiCad, nickel metal hydride, carbon zinc
(heavy duty), lithium or any rechargeable batteries. They
will not power the pump properly, and the battery life indi-
cator on the home screen may not show the correct amount.
☞
Use only Deltec Cozmo
™
3mL Insulin Cartridges; other
manufacturers’ products will not work with the Deltec pump.
☞
Strong household cleaners and chemicals will damage
the pump’s outer shell if applied directly or indirectly (on
your hands, for instance). Avoid the following in particular:
• Household cleaners (such as Formula 409
®
)
• Jewelry cleaner
• Petroleum-based products (such as gasoline or GooGone
®
)
• Products that contain high levels of D.E.E.T. (such as
Deepwoods Off
®
insect repellent)
☞
Never use abrasive cleaners, solvents, bleach, scouring
pads or sharp instruments when cleaning your pump,
as they can scratch, discolor or damage the pump’s outer
shell. If the display is scratched, it may be difficult to read
and you will need to have it replaced. If the outer shell is
chipped or cracked, it will no longer be watertight and will
require service.
☞
Never use steam or very hot water (exceeding 120˚F
[49˚C]) in an attempt to sterilize the pump. Never put
your pump in the dishwasher. Exposing the pump to these
high temperatures could damage the pump’s electronics and
result in the need to service your pump.