Deltec Cozmo Uživatelský manuál Strana 44

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 137
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 43
Using the Pump - Basics
36
2. Attaching an infusion set and loading a filled car-
tridge into the pump
Warning: To avoid accidentally infusing insulin,
disconnect the infusion set tubing from the access
device (or site) before removing a used cartridge
or replacing an infusion set. Never use the pump’s
“load cartridge” or “fill tubing” features while an
infusion set is connected to your body. You could
deliver an unintended dose of insulin, resulting in
low blood glucose (hypoglycemia).
1. Turn the cartridge cap counter-
clockwise (left) approximately
1
/
4
turn and remove it from the
pump (see Figure 9). If re-
quired, remove the used
cartridge (you may get the
“Cartridge Removed” alert;
press OK). NOTE: turning the
cartridge cap may also turn the
cartridge, so it may already be
disconnected from the pushrod when the cap is
removed. If not, turn the cartridge
1
/
4
turn to the left
to disconnect it from the pushrod.
2. Place the cartridge cap over the tip of the new car-
tridge; remove the protective
cap from the luer end of the
infusion set and insert the luer
through the hole in the cap and
firmly tighten it onto the tip of
the cartridge (see Figure 10).
Warning: If not properly
tightened, insulin could leak
from this connection and
disrupt delivery (the insulin
that leaks out is not being delivered into your body
like it should be and, depending on how much
insulin is leaking, could cause your BG level to rise).
Signs of leakage can also mean opportunity for
contamination leading to infec-
tion.
3. Insert the new cartridge into the
cartridge chamber (see Figure
11). Gently turn the cartridge
clockwise (right) about
1
/
4
turn
to fasten it onto the pushrod
(you may need to first turn the
cartridge until it drops onto the
pushrod).
D. Zurn, 2002-10-08
«cartridge-with-cover.eps»
Figure 10
Figure 11
D. Zurn, 2002-10-10
«7B-cover-insert-cartrid
g
Figure 9
D. Zurn, 2002-10-18
«9-remove-cartridge-c
o
Zobrazit stránku 43